Chatillon® TLC Series
Interchangeable Loadcell Sensors

The CHATILLON® TLC Series loadcells are full-bridge strain gauge sensors complete with 15-pin plug. These loadcells are used exclusively with the all-new Chatillon TCD Series digital force testers. They may also be used with the Chatillon DFS-R-ND digital indicator (recalibration with the DFS-R-ND is required). The loadcell values and sensitivity factors are stored in memory within the sensor plug. When connected to a TCD Series tester or DFS-R-ND indicator, the capacity and calibration points are automatically recognized by the instrument. The instrument then becomes a full-featured load measurement instrument. The loadcells are connected to the TLC Series tester or DFS-R-ND with a retractable cord that extends fully to nearly 9 ft. (3 m).
Model |
ozf |
gf |
kgf |
N |
Deflection |
Accuracy |
TLC-250G |
8 x 0.002 |
250 x 0.05 |
0.2 x 0.0001 |
2.5 x 0.0005 |
0.013 +/- 0.004 |
+/- 0.1%
|
TLC-0002 |
32 x 0.005 |
1000 x 0.1 |
1 x 0.0001 |
10 x 0.001 |
0.013 +/- 0.004 |
+/- 0.1%
|
TLC-0010 |
160 x 0.02 |
5000 x 0.5 |
5 x 0.0005 |
50 x 0.005 |
0.013 +/- 0.004 |
+/- 0.1%
|
TLC-0025 |
400 x 0.05 |
10,000 x 1 |
10 x 0.001 |
100 x 0.01 |
0.015 +/- 0.004 |
+/- 0.1%
|
TLC-0050 |
800 x 0.1 |
25,000 x 2 |
25 x 0.002 |
250 x 0.02 |
0.006 +/- 0.003 |
+/- 0.1%
|
TLC-0100 |
1600 x 0.2 |
50,000 x 5 |
50 x 0.005 |
500 x 0.05 |
0.008 +/- 0.003 |
+/- 0.1%
|
TLC-0200 |
- |
- |
100 x 0.01 |
1000 x 0.1 |
0.008 +/- 0.003 |
+/- 0.1%
|
TLC-0500 |
- |
- |
250 x 0.02 |
2000 x 0.2 |
0.008 +/- 0.003 |
+/- 0.1%
|
TLC-1000 |
- |
- |
500 x 0.05 |
5000 x 0.5 |
0.012 +/- 0.003 |
+/- 0.1%
|
TLC-2000 |
- |
- |
1000 x 0.1 |
10000 x 1 |
+/- 0.1%
|
|
TLC-5000 |
- |
- |
2500 x 0.2 |
20000 x 2 |
+/- 0.1%
|
|
TLC-10000 |
- |
- |
5000 x 0.5 |
50000 x 5 |
+/- 0.1%
|
Specifications
Input Resistance: 1000 ohms +100
Output Resistance: 1000 ohms +100
Rated Output: 2.0 mV/V
Safe Overload: 150% Full Scale
Non-Linearity: +0.05% Full Scale
Hysteresis: +0.05% Full Scale
Zero Balance: +1.0% Full Scale
Operating Temp: 30° to 120°F (-40° to 82°C)
Dimensions - TLC Series